Products/ Portable Gas Monitors

Special Features
Internal data logger (EEPROM)
Real time clock
Digital & Analog Outputs
Three 0-2.5 VDC analog inputs for logging other measurements, such as external temperature,
Pressure and humidity
Selectable data averaging times of 10 s, 1 min, 5 min and 1 hr
DewLine,M for elimination of any water vapor interference
The Model 400 Nitric Oxide Monitor is designed for the measurement of atmospheric nitric the range of 2-2000 ppbv (2 ppmv).lt is based on Patented NOZONE Technology. oxide (NO) in
A Small Concentration of OZONE is added to the sample stream and the resulting decrease of OZON is measured by the absolute method of UV absorption. in concentration
Model 400
Specifications
Measurement Principle
Titration of NO with Ozone with Detection of Ozone Depletion by UV Absorption at 254 nm
Outputs
RS-232, Flash Card Option
Precision and Accuracy
Higher of 2 ppbv or 3% of reading
Calibration
No Cal Gas Required, Annual Calibration
Data Storage
14,336 lines (10 s avg. = 1.4 days; 5 min avg = 1.4 mo.)
Time/Measurement
10 s (Data averaging options: 10 s, 1 min, 5 min, 1 hr)
Data Outputs
RS-232, LCD, Flash Card Option
Power Requirements
12 V dc or 120/240 V ac, 11 watt
Size
3.5 x 8.3 x 11.6 in. (9 x 21 x 30 cm)
Weight
2.9 kg Approximately