Using these gases it may be possible to cross-calibrate sensors for other gases and using a beakers of different volumes rating. It is possible of to Calibration at different ranges than the concentration mentioned on the Ampule.
Before use, it is important that both the ampoules and the calibration beaker are clean and dry. Ampoules should not be
used if the glass is cracked orfractured and the calibration beaker should not be used unless the lid is tight fitting.
Because some gases (e.g. Cl2) are easily absorbed it is recommended that the internal surfaces of the calibration beaker
Be permeated with the calibration gas prior to actual calibration. This is most easily accomplished by placing an ampoule
in the beaker, placing the beaker over the sensor and breaking the ampoule as if performing a normal calibration routine.
The ampoule should be left in place for approximately 5 minutes.
Model 84
Specifications
Gas/Gas Code
Ranges
Ranges
Ranges
SO2 / SD
20 PPM
50 PPM
100 PPM
H2S / HS
20 PPM
50 PPM
100 PPM
CO / CO
5 PPM
100 PPM
250 PPM